MAP Technology

9 - 8

PN 89-00002-00-063 Rev. A

One-Step

Carbodiimide

Coupling of

Oligonucleotides to

xMAP Carboxylated

Microspheres

Introduction

These protocols are a general starting point for developing assays.
All assays should be optimized for your reagents in your specific
application. Methods that give you the best results start with these
guidelines and are modified for your specific needs.

Equipment

•

Vortex

•

Sonicator bath

•

Micropipetters: (1 µl - 1000 µl)

•

Microcentrifuge

•

Timer

•

Analytical balance

Materials

•

xMAP carboxylated microspheres—LIMIT EXPOSURE TO
LIGHT!

•

Amino-substituted oligonucleotide (See Technical note 1)

•

Microcentrifuge tubes: 1.5 ml, polypropylene

•

Pipetter tips: 1 µl - 1000 µl

•

EDC:
1-Ethyl-3-(3-dimethylaminopropyl) carbodiimide hydrochloride,
Pierce Chemicals

•

MES:
0.1 M 2- (N-morpholino) ethane sulfonic acid pH 4.5, filter and
store at 4°C.

•

Tween

20 (0.02% v/v)

•

SDS:
Sodium Dodecyl Sulfate (0.1% w/v)

Preparation

1. Allow all reagents to warm to room temperature.

2. Resuspend the amino-substituted oligonucleotide to 1.0 mM in

sterile, deionized water.