Clean Sample Probe

‹

Clean the sample probe on a monthly basis, or whenever you

suspect that it may be clogged.

1. Remove the sample probe as follows: Unsnap the light housing

located above the probe. Unscrew the Cheminert fitting that
holds the sample probe in place. Gently grasp the probe and push
up. Remove the sample probe out of the top of the sample arm.

2. Use a bath sonicator or 10 mL syringe to clean the sample probe.

If you are using a sonicator, place the smaller end of the sample
probe in the sonicator for 2-5 minutes. If you are using a syringe,
force 10%-20% bleach through the larger end of the sample
probe.

3. Replace the sample probe. Adjust the vertical height if necessary.

4. Perform the following commands:

• Three Backflushes
• Three Drain commands
• Two Alcohol Flushes using 70% isopropanol or 70% ethanol.
• Three Washes - use water.

Figure 7-1. Sample Probe

Replace Air Intake

Filters

You should replace the filters on the Luminex 100 analyzer and
Luminex XYP every six months.

‹

To replace the air intake filter on the Luminex 100 analyzer:

1. Turn off the power switch on the rear of the Luminex 100

analyzer, then unplug the power cord from the wall source.

Note:

The sample probe should

slide up easily while removing it
from the sample arm. If you feel
resistance, do not force the
probe up. Contact Technical
Support.

Warning:

Be sure that the

system is not operating when you
remove the sample probe.

Vertical Adjustment Screw

Sample Probe

Note:

Hold on to the tubing! Do

not allow the tubing to fall inside
the analyzer.