Analyzer menu commands – Luminex 100 User Manual Version 1.7 User Manual

Page 49

Advertising
background image

x

MAP Technology

System Startup

PN 89-00002-00-063 Rev. A

5 - 9

Figure 5-11. Luminex Calibration Detail Report

The calibration detail report is often used by Luminex field service
personnel when performing system maintenance. The report shows
system temperature, Voltage, and pressure readings.

Analyzer menu

commands

Use the Analyzer commands to prepare the Luminex 100 analyzer
for sample acquisition, analysis, and shut down. On the Analyzer
menu, choose the command you want to perform. Many of these
commands also have buttons on the main screen.

Calibrate: Opens the Start Calibration dialog box. See page 5-2 for
calibration procedures.

Setup: Opens the Machine Setup dialog box. System connections
must be defined with initial installation of the instrument, and again
if a Luminex XYP instrument is added to the system. See page 4-3
for more information.

Sanitize: Decontaminates the sample lines and cuvette after
biohazard contact. Use a 10%-20% bleach solution to sanitize the
system after running biohazardous samples. See page 7-3 for the
sanitizing procedure.