Installation

Luminex 100

System Installation

Ensure that the facility complies with all system and safety
requirements.

If you are using the optional Luminex SD, install it after completely
installing the Luminex 100 analyzer and the Luminex XYP. See the
Luminex Sheath Delivery System appendix for instructions.

‹

To set up and start up the Luminex 100 system:

1. Unpack the Luminex 100 analyzer, Luminex XYP instrument,

and the PC. Review the hardware list on page 3-1 and identify
each system component.

2. Place the Luminex XYP on a clean, flat surface.

3. Remove the red shipping pin. Leave the silver knob in place.

Figure 4-1. XYP With Shipping Pin in Place

4. Attach the power cord and the communication cable to the

Luminex XYP. Do not plug the power cord into the outlet.

5. Place the Luminex 100 on top of the Luminex XYP. Do not plug

it into the power outlet.

4

Note:

The Luminex 100 System is

too heavy for one person to lift
alone. Two people should unpack
and lift the system to the
workbench.