Add new reporter channel calibration lots – Luminex 100 User Manual Version 1.7 User Manual

Page 44

Advertising
background image

Luminex 100 User Manual Version 1.7

x

MAP Technology

5 - 4

PN 89-00002-00-063 Rev. A

Add New Reporter
Channel Calibration Lots

Before you perform initial calibration for the analyzer, and whenever
you receive a new supply of xMAP Calibration microspheres, enter
the product and lot numbers and the calibration target values.

‹

To add a new lot of xMAP reporter calibration microspheres:

1. Click Calibrate. The Start Calibration dialog box opens.

Figure 5-4. Start Calibration Dialog Box

2. Click New in the reporter channel area. The Add New Reporter

Calibrator dialog box opens.

Figure 5-5. Add New Reporter Calibrator Dialog Box

3. Fill in the information on this dialog box using the values on the

Certificate of Quality Analysis included with your xMAP
calibration microspheres. Press Enter.

4. Click OK. You can now use this xMAP reporter channel

calibration microsphere lot to calibrate your instrument.