Safety considerations – ZOLL M Series Defibrillator Rev YH User Manual

Page 16

Advertising
background image

M S

ERIES

O

PERATOR

S

G

UIDE

1-8

Safety Considerations

The M Series products are high energy devices capable
of delivering up to 360 joules. To completely deactivate
the device, you must turn the SELECTOR SWITCH to
the OFF position.

In order to disarm a charged defibrillator:

Turn the SELECTOR SWITCH to MONITOR, OFF or
PACER (pacer equipped versions only)
or

Change the selected defibrillator energy

As a safety feature, the device will automatically disarm if
left charged for more than 60 seconds (15 seconds for
AED versions).

Note: The terms “ZOLL Multi-Function Electrode (MFE)

Pads” and “MFE Pads” will be used
interchangeably throughout this manual.

WARNINGS

General

Federal (U.S.A.) law restricts this device to use by or on
the order of a physician.

The use of external pacing/defibrillation electrodes or
adapter devices from sources other than ZOLL is not
recommended. ZOLL makes no representations or
warranties regarding the performance or effectiveness of
its products when used in conjunction with pacing/
defibrillation electrodes or adapter devices from other
sources. Device failures attributable to the use of pacing/
defibrillation electrodes or adapters not manufactured by
ZOLL may void ZOLL's warranty.

Proper operation of the unit, together with correct
electrode placement is critical to obtaining optimal
results. Operators must be thoroughly familiar with
proper device operation.

Do not use the unit in semiautomatic mode during patient
movement. A patient must be motionless during ECG
analysis. Do not touch the patient during analysis. Cease
all movement via stretcher or vehicle before analyzing
the ECG. If using the device in an emergency vehicle,
bring the vehicle to a halt before using in semiautomatic
mode.

Place the patient on a firm surface before performing
CPR.

The device is protected against interference from radio
frequency emissions typical of two-way radios and
cellular phones (digital and analog) used in emergency
service/public safety activities. Users should assess the
device’s performance in their typical environment of use
for the possibility of radio frequency interference from
high-power sources. Radio Frequency Interference (RFI)
may be observed as shifts in monitor baseline, trace
compression, display brightness changes or transient
spikes on the display.

M Series units equipped with the Bluetooth® option
include an RF transmitter which transmits with 0dBm
power in the 2.4 GHz ISM band.

Do not operate the unit without a battery. Keep a fully
charged spare battery pack with the device at all times.

Regular use of partially charged battery packs without
fully recharging between uses will result in permanently
reduced capacity and early battery pack failure.

Test batteries regularly. Batteries that do not pass ZOLL’s
capacity test could unexpectedly shutdown without
warning.

Replace the battery with a fully charged battery
immediately after the “LOW BATTERY” or “REPLACE
BATTERY” message.

Emergency defibrillation should be attempted only by
appropriately trained, skilled personnel who are familiar
with equipment operation. Training appropriateness, such
as Advanced Cardiac Life Support (ACLS) or Basic Life
Support (BLS) certification, should be determined by the
prescribing physician.

Synchronized cardioversion should only be attempted by
skilled personnel trained in Advanced Cardiac Life
Support (ACLS) and familiar with equipment operation.
The precise cardiac arrhythmia must be determined
before attempting defibrillation.

Prior to attempting synchronized cardioversion, ensure
that the ECG signal quality is good and that sync marks
are displayed above each QRS complex.

These operating instructions describe the functions and
proper operation of the M Series products. They are not
intended as a substitute for a formal training course.
Operators should obtain formal training from an
appropriate authority prior to using the device for patient
care.

Do not disassemble the unit. A shock hazard exists. Refer
all problems to authorized service personnel.

Follow all recommended maintenance instructions. If a
problem occurs, obtain service immediately. Do not use
the device until it has been inspected by the appropriate
personnel.

Do not use the unit’s ECG out signal as a sync pulse for
another defibrillator or cardioverter.

Advertising