Pacer (Pacer Version Only)

Symptom

Recommended Action

15. “CHECK PADS” message.

Ensure MFE Pads are connected to Multi-Function cable.

Ensure electrode gel is not dry. Replace MFE Pads if
necessary.

Ensure good electrode-to-patient contact.

Check integrity of Multi-Function cable by plugging into test
connector. “CHECK PADS” should disappear.

16. No stimulus marker (

) present on the ECG

trace.

Ensure unit is in PACER mode.

Ensure PACER RATE (ppm) is set greater than patient heart
rate.

17. No ventricular capture beat after stimulus marker on

ECG display.

Check patient’s pulse.

Increase output current.

Ensure MFE Pads are making good contact with the patient.

Select different ECG Lead configuration.

Review MFE Pad placement.

18. Patient on “Standby” pacing gets paced

intermittently.

Ensure proper ECG electrode connection and placement. If
ECG lead wire comes off, pacer will automatically pace
asynchronously.

Check ECG cable for damage.

Patient R-to-R interval varying. Pace rate close to patient rate.

Verify rate is set appropriately

19. Heart rate display reads 0 with proper pacing

capture displayed on ECG trace.

Check patient’s pulse.

Select different ECG Lead configuration.

20. Bedside/Central Station/Telemetry ECG display

becomes erratic when pacing.

Patient monitor ECG inputs overloaded by pacer signals.
ECG can only be monitored by the device while pacing.