C-4

Display Messages

Message

Reason/Action

Card Removed

Card was removed from
PCMCIA slot or is not seated
properly.

Reinsert card.

Empty Card

Empty data card inserted into
PCMCIA slot or card was
erased.

Record data on card before
transmission or insert different
card.

Not a Data Card

Wrong type of card.

Insert a data card.

Host Fault

Transmission was stopped
because PC detected an error
or user cancelled transmission
via the PC communication
program.

Check the communications
package or utility running on
the PC for the source of the
error.

Serial Fault

The cable is not connected, the
PC host is not detected, or a
transmission error occurred.

Check and secure the RS 232
cable at the back of the unit or
the PC. Check the settings on
the communication package or
HyperTerminal.

Card Uploaded

Upload successful

Card Previously
Uploaded

User inserted a valid data card
that was already uploaded
successfully. Data on card not
erased.

Use a different data card, erase
the data on the card you are
using, or resend the data
again.