Electromagnetic Immunity Declaration (EID)

Guidance and manufacturer’s declaration — electromagnetic immunity for the M Series.

The M Series is intended for use in the electromagnetic environment specified below. The customer or user of the
M Series should ensure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment –
guidance

Electrostatic
discharge (ESD)
IEC 61000-4-2

±6 kV contact

±8 kV air

±6kV contact

±8 kV air

Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.

Electrical fast
transient/burst
IEC 61000-4-4

±2 kV for power supply
lines

±1 kV for input/output lines

±2 kV for power supply
lines

Not applicable

Mains power quality should be that of a
typical commercial or hospital
environment.

Surge
IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

±1 kV differential mode

±2 kV common mode

Mains power quality should be that of a
typical commercial or hospital
environment.

Voltage dips, short
interruptions and
voltage variations
on power supply
input lines.
IEC 61000-4-11

<5% U

T

(>95% dip in U

T

)

for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in U

T

)

for 5 seconds

<5% U

T

(>95% dip in

U

T

) for 0.5 cycle

40% U

T

(60% dip in

U

T

) for 5 cycles

70% U

T

(30% dip in

U

T

) for 25 cycles

<5% U

T

(>95% dip in

U

T

) for 5 seconds

Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the M Series
requires continued operation during
power mains interruptions, it is
recommended that the M Series be
powered by an uninterruptible power
supply or a battery.

Power frequency
(50/60 Hz)
magnetic field.
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should
be at levels characteristic of typical
location in a typical commercial or
hospital environment.

NOTE: U

T

is the ac mains voltage prior to the application of the test level.