Recorder
PCMCIA Card
Battery Packs
Paper
80 mm thermal (grid width)
90 mm (paper width)
Speed
25 mm/sec
Delay
6 seconds
Annotations
Time, date, defib energy, heat rate, pacer output (Pacer version only), QRS sync marker,
ECG size, lead, alarm, defib test OK/Fail, analyze ECG, Pads off, analysis halted, noisy ECG,
shock advised, no shock advised, ECG too large, and diagnostic bandwidth.
Printing Method
High resolution, thermal array printhead.
Print-out Modes
Manual or automatic — user-configurable.
On/Off Control
Front panel and paddle.
Automatic Function
15 second recording initiated by alarm activation or defibrillator discharge.
Capacity
Standard series II flash card —1 megabyte to 16 megabytes.
Audio Recording
Digital compressed audio data.
PD 4410
XL Battery
Type
Rechargeable, sealed lead acid
Rechargeable, sealed lead acid
Weight
1 kg (2.2 lbs)
1.7 kg (3.7 lbs)
Voltage
2V/cell; 5 cells wired in series
2V/cell; 5 cells wired in series
Recharge Time
4 hours or less with integral charger
7.2 hours or less with integral charger
Operating Time
For a new, fully charged battery pack at
20 °C: 35 defibrillator discharges at
maximum energy (360 J), or 2.75 hours of
continuous ECG monitoring, or 2.25 hours of
continuous ECG monitoring/pacing at
60 mA, 80 beats/min.
For a new, fully charged battery pack at
20 °C: 60 defibrillator discharges at
maximum energy (360 J), or 4 hours of
continuous ECG monitoring, or 3.25 hours of
continuous ECG monitoring/pacing at
60 mA, 80 beats/min.
Note: Each monitoring option added to the M Series device decreases the Operating Time
that can be obtained from a fully charged battery. Refer to the individual Option Insert for
the Operating Run Time specific to your device. For further details specific to your
device, contact the ZOLL Technical Service Department.
Low Battery
Indicator
Message displayed on monitor and 2-beep low battery tone sounds once a minute until just
before shutdown, when it will beep twice every 2 seconds. The time from display of the “LOW
BATTERY” or “REPLACE BATTERY” message until the instrument shuts down will vary
depending upon the battery age and condition.