Operator’s Shift Checklist for M Series Products (Semi-Automatic)

Recommended checks and procedures to be performed at
the start of each shift. For more detailed information, see
the M Series Operator’s Guide.

Date

______________________________________________

Location ____________________________________________

Unit Serial Number ___________________________________

1. Condition

1st

Shift

2nd

Shift

3rd

Shift

Remarks

Unit clean, no spills, clear of objects on top, case intact

2. Multi-function Pads

1 set pre-connected / 1 spare

3. Paddles (if applicable)

Paddles clean, not pitted

Release from housing easily

4. Inspect cables for cracks, broken wires, connector

A ECG electrode cable, connector

B Defibrillator paddle cables

C Multi-function cable, connector

5. Batteries

A Fully charged battery in unit

B Fully charged spare battery available

6. Disposable supplies

A Electrode gel or gel patches

B MFE Pads in sealed pouches — 2 sets

C ECG electrodes

D Recorder paper

E Alcohol wipes

F Razors

7. Operational checks

A Power On Sequence

Turn unit to ON, 4-beep tone heard

B Defibrillator

Multi-function cable connected to test connector: “CHECK PADS” displays.

Press ANALYZE button, unit charges to 30 J

Press and hold SHOCK button, “TEST OK” is displayed and printed

Attach MFC to ECG Simulator, set to VF

Verify “CHECK PATIENT” message is displayed

Press ANALYZE. Verify unit charges to 200 J

Press SHOCK, verify shock was delivered.

C Paddles (if applicable) (Manual Mode ONLY)

Paddles in holder: Set defib energy level to 30 joules, press paddles firmly into

the side wells, and simultaneously press and hold both defib discharge buttons;

“TEST OK” message on Recorder.

D Pacer Operation (Manual Mode ONLY)

Multi-function cable not connected to Test Connector

Turn to PACER, set pacer rate to 150 ppm, press RECORDER button

Pacer pulses occur ever 2 large divisions (10 small divisions)

Press 4:1 button, pulses occur every 8 large divisions

Set PACER OUTPUT to 0 mA, no “CHECK PADS” message

Set PACER OUTPUT to 16 mA, “CHECK PADS” message and alarm

Reconnect Multi-function cable to test connector.

Press Clear Pace Alarm softkey; “CHECK PADS” message disappears and

Pace alarm stops.

E Recorder

Press RECORDER button; Recorder runs. Press again; Recorder stops.

Inspect Recorder printing

8. Please check the appropriate box after each use of this checklist.

Signatures

No action required

1st ___________________

Minor problem(s) corrected

Disposable supplies replaced

2nd __________________

Major problem(s) identified — UNIT OUT OF SERVICE

3rd ___________________