ZOLL M Series Defibrillator Rev YH User Manual
Page 69

10-7
Operator’s Shift Checklist for M Series Products (Semi-Automatic)
Recommended checks and procedures to be performed at
the start of each shift. For more detailed information, see
the M Series Operator’s Guide.
Date
______________________________________________
Location ____________________________________________
Unit Serial Number ___________________________________
1. Condition
1st
Shift
2nd
Shift
3rd
Shift
Remarks
Unit clean, no spills, clear of objects on top, case intact
2. Multi-function Pads
1 set pre-connected / 1 spare
3. Paddles (if applicable)
Paddles clean, not pitted
Release from housing easily
4. Inspect cables for cracks, broken wires, connector
A ECG electrode cable, connector
B Defibrillator paddle cables
C Multi-function cable, connector
5. Batteries
A Fully charged battery in unit
B Fully charged spare battery available
6. Disposable supplies
A Electrode gel or gel patches
B MFE Pads in sealed pouches — 2 sets
C ECG electrodes
D Recorder paper
E Alcohol wipes
F Razors
7. Operational checks
A Power On Sequence
Turn unit to ON, 4-beep tone heard
B Defibrillator
Multi-function cable connected to test connector: “CHECK PADS” displays.
Press ANALYZE button, unit charges to 30 J
Press and hold SHOCK button, “TEST OK” is displayed and printed
Attach MFC to ECG Simulator, set to VF
Verify “CHECK PATIENT” message is displayed
Press ANALYZE. Verify unit charges to 200 J
Press SHOCK, verify shock was delivered.
C Paddles (if applicable) (Manual Mode ONLY)
Paddles in holder: Set defib energy level to 30 joules, press paddles firmly into
the side wells, and simultaneously press and hold both defib discharge buttons;
“TEST OK” message on Recorder.
D Pacer Operation (Manual Mode ONLY)
Multi-function cable not connected to Test Connector
Turn to PACER, set pacer rate to 150 ppm, press RECORDER button
Pacer pulses occur ever 2 large divisions (10 small divisions)
Press 4:1 button, pulses occur every 8 large divisions
Set PACER OUTPUT to 0 mA, no “CHECK PADS” message
Set PACER OUTPUT to 16 mA, “CHECK PADS” message and alarm
Reconnect Multi-function cable to test connector.
Press Clear Pace Alarm softkey; “CHECK PADS” message disappears and
Pace alarm stops.
E Recorder
Press RECORDER button; Recorder runs. Press again; Recorder stops.
Inspect Recorder printing
8. Please check the appropriate box after each use of this checklist.
Signatures
No action required
1st ___________________
Minor problem(s) corrected
Disposable supplies replaced
2nd __________________
Major problem(s) identified — UNIT OUT OF SERVICE
3rd ___________________