ZOLL M Series Defibrillator Rev YH User Manual

Page 86

Advertising
background image

M S

ERIES

O

PERATOR

S

G

UIDE

B-2

5. “POOR LEAD CONTACT” message.

Check that ECG cable is connected to patient and instrument.

Check that ECG electrodes are making good contact and not
dried out.

If changing from five (5) lead ECG patient cable to three (3)
lead ECG patient cable, turn unit OFF for at least 10 seconds.

Replace ECG cable.

6. Noisy ECG, Artifact, Wandering Baseline.

Consider 1 – 21Hz filter bandwidth (see M Series
Configuration Guide)

Prepare the patient’s skin prior to electrode attachment.

Check for proper adhesion of electrodes to patient.

Reduce or eliminate ECG artifact due to electrode or patient
cable movement. Route cables so that they don’t pull on
electrodes or swing excessively.

Ensure patient is motionless.

Check for possible excessive radio frequency interference.

7. Poor ECG signal level, calibration pulse normal.

Select another lead.

Apply new electrodes using different placement.

8. Inconsistent QRS beep or heart rate.

Increase beeper volume.

Select another lead.

Alter ECG electrode placement.

9. Sync marker is absent or inconsistent with QRS

waveform on display and recorder.

Ensure device is in SYNC mode.

Change ECG lead selection.

Alter ECG electrode placement.

Paper too narrow. It should be 90 mm wide.

10.

Unit displays the CABLE FAULT or PADDLE
FAULT
message.

Check the connection between the Multifunction cable and
the MFE pads.

Remove the CPRD-to-MFC connector, if in use, and plug the
multifunction cable directly into the CPR-Dpadz.

Symptom

Recommended Action