ZOLL M Series Defibrillator Rev YH User Manual
Page 28

M S
ERIES
O
PERATOR
’
S
G
UIDE
2-6
Analyze Format
The summary report records 6 seconds of pre analysis ECG and 9 seconds of ECG recorded during the ECG analysis
interval with the annotation “SHOCK ADVISED” or “NO SHOCK ADVISED”. AED units will additionally include shock
count and AED mode annotations.
Manual Mode Activated
AED versions of the M Series that are equipped with manual mode capabilities will record a “MANUAL MODE
STARTED” event within Summary Report whenever the device is switched from the AED (Default) mode to the Manual
operating mode.
The following annotations may also be printed at the top of the Analyze Format printout:
Annotation
Description
1. POOR PAD CONTACT:
The MFE Pads are detected as having poor connection.
2. ANALYSIS HALTED:
The ECG analysis is halted due to either the ANALYZE button being pushed or a fault
condition.
3. NOISY ECG:
Excessive noise is detected.
4. SHOCK ADVISED:
Shockable rhythm has been detected at the end of user-initiated ECG analysis.
5. NO SHOCK ADVISED
No shockable rhythm has been detected at the end of user-initiated ECG analysis.
6. ECG TOO LARGE
The amplitude of the ECG signal is too large for proper rhythm analysis.
DEFIB ADVISORY
ECG LEAD
ELECTRODES I
I
NAME
ECG SIZE
2
09:56:26 10 FEB 98 NO SHOCK ADVISED
NOISE EVENTS
0
NO SHOCK ADVISED