Troubleshooting Guides
B-3
Recorder
Symptom
Recommended Action
11. “CHECK RECORDER” message.
•
Recorder out of paper.
•
Remove paper, check paper type, check recorder for paper
jam, reload paper.
•
Recorder door is open.
12. Recorder makes stuttering sound when activated.
•
Check recorder for paper jam.
13. Light or poor quality printing.
•
Ensure correct paper is in use.
•
Ensure paper is installed grid side against recorder print head.
•
Recorder print head requires cleaning (trained personnel
only).
14. Summary Report will not print when SUMMARY
button is depressed.
•
15 seconds have not elapsed since one of the events that
trigger Summary Report to record have occurred. Wait 15
seconds and try again.