Operating Controls and Indicators

2-7

Printing a Report

To print the stored information, press the SUMMARY
button below the screen display. Then press the
corresponding softkey to print configuration settings,
print chart, or print log.

The recorder will print all events currently in memory in
chronological order. If the stripchart recorder is on or the
defibrillator is charged, summary report printing is
disabled. To stop printing a report, press the
RECORDER button or turn the unit off. An unlimited
number of copies of the report may be printed by simply
pressing the SUMMARY button and corresponding print
softkey again.

Pressing the RECORDER button while printing a
Summary Report will cause the unit to stop printing the
report. Press the RECORDER button again to begin
printing an ECG trace. The stripchart recorder will run
continuously until the button is pressed again.

Pressing the SUMMARY button and a corresponding
print softkey while printing a report will cause the current
report to stop printing and a new report to begin printing.

Printing is interrupted if a vital sign alarm occurs (i.e. HR,
SpO

2

, etc.), the ANALYZE button is pressed, or the

defibrillator is charged.

If the recorder is out of paper when the SUMMARY
button and a corresponding print softkey are pressed, a
“CHECK RECORDER” message will appear on the
display. Load paper and press SUMMARY again to
select the report to print.

Printing Part of a Report

If you want to print out only a portion of the Summary
Report:

1. Press the SUMMARY button.
2. Press the Print Chart softkey.
3. Press the Print Range softkey.
4. Press Prev. Event or Next Event softkey to scroll

through the events.

5. Press Print softkey.

The M Series unit prints the displayed event and all
following events.

Adding Patient Name and ID# to a
Report

To add patient name and identification number to the
summary report:

1. Press ID # softkey.
2. Press Prev. Digit or Next Digit softkey to select letter

for patient name.

3. Press Inc Digit or Dec. Digit softkey to change value

of letter.

Repeat steps 2 and 3 until you have entered the patient’s
entire name.

4. Press Enter Name softkey.
5. Press Prev. Digit or Next Digit softkey to select digit

or letter for identification number.

6. Press Inc Digit or Dec. Digit softkey to change value

of digit.

Repeat steps 5 and 6 until you have entered the patient’s
entire identification number.

7. Press Enter ID and Return softkey.

Note: Patient name cannot be retrospectively added to

summary report events already stored in memory.
The patient name is only stored with summary
events saved after the patient name has been
entered.

Printing an Incident Log

An incident log is an abbreviated list of all major events
recorded in the summary report. You can print out an
incident log that includes the time of occurrence of the
following events:

M Series unit powered on.

Defibrillation advisory messages (for example, CHECK
PATIENT and SHOCK ADVISED).

Defibrillation shocks (including energy level).

Pacer mode activated.

Manual mode started (AED only).

Alarms triggered.

Code markers.

12 Lead analysis initiated.

12 Lead data transmission.

Recorder turned on.

NIBP measurements activated.

In addition, the incident log lists the following:

Report start time (time when summary report memory was
erased).

Last event time (time of last event in memory).

Total number of shocks.

Total pacer time.

System serial number.

Device identification number.

CHARGER ON

SUMMARY

CODE

MARKER