Operating controls and indicators, Selector switch, Defib energy select buttons – ZOLL M Series Defibrillator Rev YH User Manual

Page 23: Defib charge, Shock

Advertising
background image

2-1

SECTION 2

OPERATING CONTROLS AND INDICATORS

1. SELECTOR SWITCH

The selector switch allows selection of the following
modes: OFF, MONITOR, DEFIB, and PACER, (Pacer
version only)

2. DEFIB ENERGY SELECT BUTTONS

Two sets of up-down arrow buttons control the
defibrillator energy level, one set located on the front
panel and the other located on the sternum paddle.

Press and hold the appropriate up (

▲)or down (▼) arrow

button until the desired energy level is indicated on the
display.

3. DEFIB CHARGE

Pressing the CHARGE button on the front panel or, if
using paddles, on the apex paddle handle, charges the
defibrillator to the selected energy level.

4. SHOCK

The SHOCK button illuminates when the defibrillator is
charged and ready. Press and hold the button to
discharge the defibrillator.

The SHOCK button is only active when using
Multi-Function Electrodes (MFE) Pads, external
autoclavable paddles, or internal defibrillation paddles
without a discharge button. The SHOCK button is not
functional when external paddles are connected to the
unit.

Each external paddle has a SHOCK button located near
the forward end of the handle. Press and hold both
buttons simultaneously to discharge the defibrillator.

SHOCK

DEFIB

MONITOR

PACER

OFF

ANALYZE

LEAD

SIZE

4:1

ALARM

SUSPEND

CODE

MARKER

SUMMARY

SUMMARY

RELEASE

CHARGER ON

RECORDER

PACER
OUTPUT
mA

PACER

RATE

ppm

CHARGE

ENERGY

SELECT

3

2

1

CO2

Hg

RR

31

x1.5 72

17

PADS

100

Sp02 %

ECG

ECG

CO2

Param

Wave 2

Alarms

LEAD

SIZE

HR

m
m

4

1

2

5

7 8 910

22

21

20

19

16

15

14

13

6

12

11

17

18

3

23

Advertising