Defib 200j sel, Analyzing ecg, No shock adv – ZOLL M Series Defibrillator Rev YH User Manual

Page 38: Shock advised

Advertising
background image

M S

ERIES

O

PERATOR

S

G

UIDE

4-2

1

Select DEFIB

The unit displays “DEFIB 200J SEL” on the monitor until
the ANALYZE button is depressed by the operator.

Energy Select

Shock number 1 is set at 200 joules, shock 2 is set at
200 joules and shock 3 and up is set at 360 joules
(default setting). If medical protocols allow, the operator
may select a different energy level using the energy

select up (

) and down (

) arrow buttons. The new

energy setting will display on the monitor.

Manually changing the energy level outside the pre-
programmed Shock 1, 2, 3 sequence and delivering a
shock disables automatic energy escalation. See the
Energy Level: Shock 1, 2, 3 section of the M Series
Configuration Guide for more details.

2

Press ANALYZE Button

Press the ANALYZE button to begin analysis of the
patient’s ECG rhythm and to detect the presence of any
shockable rhythms.

An “ANALYZING ECG” message will be displayed for
approximately 9 to 12 seconds while the patient’s ECG is
analyzed.

Once the analysis is completed, the unit indicates
whether or not a shock is advised.

When a non-shockable rhythm is detected the message
“NO SHOCK ADV.” will be displayed.

Follow local protocols to continue CPR or other
cardiopulmonary life support and re-analyze the ECG at
appropriate intervals.

When a shockable rhythm is detected (Ventricular
Fibrillation or Tachycardia with heart rate > 150), units
with the automatic charge option enabled will
automatically charge the defibrillator to the
pre-configured or user selected energy setting.

Units with the automatic charge option disabled will
alternately display the messages “SHOCK ADVISED”
and “PRESS CHARGE” when a shockable rhythm is
detected.

WARNING

A patient must be motionless during ECG analysis. Do not
touch the patient during analysis. Cease all movement via
stretcher or vehicle before analyzing the ECG.

PACER

OFF

MONITOR

DEFIB

1

DEFIB 200J SEL.

ID#

Param

00:01

ECG

Wave 2

Alarms

Sync

On/Off

ANALYZING ECG

ID#

Param

00:01

ECG

Wave 2

Alarms

Sync

On/Off

DEFIB 200J SEL.

ID#

Param

00:01

ECG

Wave 2

Alarms

Sync

On/Off

NO SHOCK ADV.

Sync

On/Off

DEFIB 200J SEL.

Alarms

ID#

Wave 2

Param

00:01

SHOCK ADVISED

ECG

Advertising