Manual Defibrillation

3-3

Using your thumbs, simultaneously press and hold both
SHOCK buttons (one on each paddle) until energy is
delivered to the patient.

Once energy is delivered, the display will simultaneously
show “XXXJ DELIVERED” and “DEFIB XXXJ SEL.” After
approximately 5 seconds the “XXXJ DELIVERED”
message will disappear and the “DEFIB XXXJ SEL”
message remains to indicate the selected energy level.

Note: If the defibrillator is not discharged within 60

seconds after reaching the selected energy level,
the unit automatically disarms itself.

During the 10 seconds prior to disarming, the charge
ready tone will beep intermittently. The charge ready
tone will then stop, the charge indicator light will go off,
and the monitor message will change to “DEFIB XXXJ
SEL.” Press the CHARGE button to recharge the unit.

Paddle Cleaning

Paddle plates and handles must be thoroughly cleaned
after each use. See the General Maintenance section
for the correct cleaning procedure.

CAUTION

Only use thumbs to depress the SHOCK buttons.
Failure to do so could result in the inadvertent
depression of the ENERGY SELECT buttons, causing
the defibrillator to disarm itself.