APPENDIX A
SPECIFICATIONS

General

Size

6.8 in. high X 10.3 in. wide X 8.2 in. deep
17.3 cm high X 26.2 cm wide X 20.8 cm -deep

Weight

11.5 lbs (5.23 kg) with Multi-Function Cable and battery; 13.5 lbs (6.14 kg) with paddles

Power

Sealed lead acid battery pack; 5 cells, 2V/cell (wired in series)

AC Power

100-120 ~ 50/60 Hz, 220-240 ~ 50 Hz, 220 VA

DC Input (Optional)

10-29 V. 130 W

Device
Classification

Class I and internally powered per IEC 601-1.
Class II and internally powered per IE 601-1. (DC Input ONLY).

Design Standards

Meets or exceeds UL 2601, AAMI DF-39, AAMI DF-2, IEC 601-2-4, IEC 601-2-25, and
IEC 601-2-27.

Patient Safety

All patient connections are electrically isolated.

Environmental

Operating Temperature: 0

C to 55 C.

Note: The M Series device may not perform to specifications when stored at the upper or

lower extreme limits of storage temperature and immediately put into use.

Storage and Shipping Temperature: -20

 to 60C

Humidity: 5 to 95% relative humidity, non-condensing
Vibration: Mil Std 810E, Minimum Integrity Test
Shock: IEC 68-2-27, 50g 6mS half sine
Operating Pressure: 594 to 1060 mBar
Material Ingress: IEC 529, IP24
Electromagnetic Compatibility (EMC): CISPR 11 Class B - Radiated and Conducted Emissions
Electromagnetic Immunity: AAMI DF-2: IEC 1000-4-3 to 20 V/m
Electrostatic Discharge: AAMI DF-2: IEC 1000-4-2
Conducted Susceptibility: IEC 1000-4-4, 1000-4-5, 1000-4-6