General Maintenance

10-3

Multi-Function Cable Setup

The Multi-Function cable should be plugged into the unit.
Make sure the Multi-Function cable is not plugged into its
test connector.

Switch unit to DEFIB and set energy to 30 joules.

The messages “CHECK PADS” and “POOR PAD
CONTACT” will be alternately displayed.

Plug the Multi-Function Cable into its test connector.

The message “DEFIB PAD SHORT” will be displayed.

3. Energy Delivery Test (Paddles/MFE Pads)

Press the CHARGE button on the front panel or on the
apex paddle handle.

Wait for the charge ready tone to sound and verify that the
energy ready value displayed on the monitor registers 30
joules (“DEFIB 30J READY”).

If paddles are installed, using your thumbs, simultaneously
press and firmly hold the SHOCK buttons (one on each
paddle) until discharge occurs.

If MFE cable and test connector are installed, press and
hold the SHOCK button on the front panel of the
defibrillator until discharge occurs.

The stripchart recorder will print a short strip indicating
“TEST OK” and energy delivered if the unit delivered
energy within specifications.

If “TEST FAILED” appears, contact appropriate technical
personnel or ZOLL Technical Service Department
immediately.

Note: During the Energy Delivery Test, unit will only

discharge when energy level is set to 30 joules.

4. Pacer Operation (Pacer Version Only)

Turn the SELECTOR SWITCH to PACER.

Turn PACER RATE control to 150 ppm.

Press the RECORDER button to generate a strip.

Verify that the pacing stimulus markers (

) occur

approximately every 10 small divisions (2 large divisions,
1 cm).

Press the 4:1 button and verify that the frequency of the
markers decrease (8 large divisions, 4 cm between each
marker).

Turn the PACER OUTPUT control to 0 mA. There
should be no “CHECK PADS” or “POOR PAD
CONTACT” messages.

Disconnect MFE Pads or paddles from the Multi-Function
cable.

Slowly turn the knob up to 16 mA. The “CHECK PADS”
and “POOR PAD CONTACT” messages are alternately
displayed on the screen. Pace alarm sounds and flashes.

Connect Multi-Function cable to test connector.

Press the Clear Pace Alarm softkey. The “CHECK
PADS” and “POOR PAD CONTACT” messages disappear
and the Pace alarm stops.

5. Recorder Check

Check for adequate supply of paper.

Press the RECORDER button. The strip recorder will run
until the RECORDER button is pressed again.

Press and hold the SIZE button for at least 2 seconds to
generate a calibration pulse. The calibration pulse will
remain on the display for as long as the SIZE button
remains depressed. In addition, the amplitude of the
calibration pulse is 1 mv independent of the SIZE setting.

Inspect the recorder waveform for uniformity and
darkness.

Inspect for uniformity of annotation characters and
completeness of words.

Check strip recorder speed by verifying that the calibration

pulse is 2.5

½ mm

wide and 10 ± 1 mm high.

LEAD

RECORDER

SIZE

ALARM

SUSPEND

PACER

RATE

ppm

PACER
OUTPUT
mA

4:1