Pacemaker (pacer version only) defibrillator – ZOLL M Series Defibrillator Rev YH User Manual
Page 76

M S
ERIES
O
PERATOR
’
S
G
UIDE
A-2
Pacemaker (Pacer Version Only)
Defibrillator
Type
VVI demand; asynchronous (fixed rate) when used without ECG leads or in ASYNC pacing
mode.
Pulse Type
Rectilinear, constant current.
Pulse Shape and
Duration
Rectilinear, 40 milliseconds ±2 milliseconds
Pulse Amplitude
Variable 0 mA to 140 mA ±5% or 5 mA, whichever is greater.
Digitally displayed on the monitor (increments or decrements by a value of 2 mA).
Pacing Rate
Variable from 30 ppm to 180 ppm ±1.5% (increments or decrements by a value of 2 ppm).
Output protection
Fully defibrillator protected and isolated
Multi-Function
Electrode (MFE)
Pads
Specifically designed adult anterior/posterior pre-gelled ZOLL MFE Pads, and Multi-Function
stat•padz packaged in pairs.
Waveform
Dampened sinusoid.
Energy Selection
Selectable at 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 75, 100, 150, 200, 300, 360 J
(Delivered into 50
load). Selected using controls on sternum paddles or unit front panel.
Charge Time
Less than 7 seconds with a new fully charged battery (first 15 charges to 360 joules).
Depleted batteries will result in a longer defibrillator charge time.
Energy Display
Monitor display indicates both selected and delivered energy.
Synchronized Mode
Synchronizes defibrillator pulse to patient’s R-wave. “SYNC” message displayed on monitor.
Marker on display and recorder paper identifies R-wave discharge point.
Charge Controls
Control on apex paddle and on device front panel.
Paddles
Standard anterior/posterior adult and pediatric. Adult paddles slide off to expose pediatric
paddles.
Multi-Function
Electrode (MFE)
Pads
Specifically designed adult anterior/posterior pre-gelled ZOLL MFE Pads, and Multi-Function
stat•padz packaged in pairs.
Built-in Defibrillator
Tester
Provides verification of the defibrillator charging and discharging without removing paddles
from storage wells, or verification of unit configured with Multi-Function Cable.
Defibrillation
Advisory
Evaluates electrode connection and patient ECG to determine if defibrillation is required.
Shockable Rhythms: Ventricular fibrillation with amplitude > 100
V and wide complex
Ventricular tachycardia with rates greater than 150 bpm. Refer to the ECG Analysis Algorithm
Accuracy section for sensitivity and specificity performance.
Multi-Function
Electrode
Impedance
Measurement
Accuracy
0 ohms - 250 ohms ±10% or 5 ohms, whichever is greater.