AC Charger
27. “CHECK PADS” message.
•
Verify proper Multi-Function Cable / MFE Pad connection by
disconnecting and reconnecting the Mult-Function cable and
MFE Pads.
•
Ensure proper contact of Multi-Function Pads and that the
patient does not have excessive hair beneath the electrodes.
•
If message persists, disconnect Multi-Function cable from
MFE Pads and plug cable into test connector. “CHECK
PADS” should change to “DEFIB PAD SHORT” (Manual
Mode Only).
•
If test fails, try using paddles to defibrillate.
28. “USE PADS” message.
•
ECG Analysis will only operate when MFE pads are attached
to the patient.
•
Disconnect paddle and connect MFE Pads for use in Semi-
automatic defibrillation.
•
Activate manual mode to use Paddles.
29. “NOISY ECG”
“RETRY ANALYSIS” message.
•
Check for proper application and adhesion of Multi-Function
Electrodes.
•
Check to make sure that nobody is touching the patient and
that the patient is motionless.
30. “ECG TOO LARGE”
“RETRY ANALYSIS” message.
•
Press ANALYZE button again to begin analysis.
31. No “TEST OK” message when performing a
defibrillator self-test.
•
Check to make sure unit is set to 30 joules.
•
If testing with Multi-Function Cable, make sure that cable is
firmly inserted into test connector.
•
If testing with paddles, make sure to press the paddles firmly
against the sides of the unit while discharging.
32. “DEFIB MAINT. REQUIRED” message.
•
Contact ZOLL Technical Service Department.
Symptom
Recommended Action
33. The green and orange-yellow CHARGER ON
indicators are alternately illuminating.
•
Verify battery is installed.
•
Turn unit ON to identify the fault condition.
•
Replace battery pack with a fully charged battery pack.
•
If problem persists, replace battery pack, unplug device from
AC mains and plug device back into AC mains.
Symptom
Recommended Action