A-5
Guidance and Manufacturer’s Declaration — Electromagnetic Emissions
The M Series is intended for use in the electromagnetic environment specified below. The customer or user of the
M Series should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF Emissions
CISPR 11
Group 1
The M Series uses RF energy for its internal function only. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
Harmonic Emissions
IEC 61000-3-2
Class A
The M Series is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies