ZOLL Propaq M Rev A User Manual
Page 12
Advertising
C
HAPTER
1
M
AINTENANCE
T
ESTS
8
9650-0804-01 Rev. A
Acceptable Results
The unit passes the test if P
d-Primary
is less than or equal to 3.0 mmHg and P
d-Safety
is less than
or equal to 4.0 mmHg for all three target pressures.
Corrective Action
If readings fail to meet the required accuracy, the NIBP module requires replacement or factory
analysis and realignment.