ZOLL Propaq M Rev A User Manual
Page 62
Advertising
C
HAPTER
4
L
IFE
C
YCLE
C
OUNTERS
58
9650-0804-01 Rev. A
CO2
On the Propaq M, do the following:
1. Press the Setup quick access key (
).
2.
Use the navigation keys to select Supervisor>Service>Diagnostic>CO2>Service Codes.
3.
Select Reset CRC Counter. Select Yes when prompted to confirm the reset.
See “Functional Verification Tests” on page 1-22 to perform the functional verification tests.
SpO2
On the Propaq M, do the following:
1. Press the Setup quick access key (
).
2.
Use the navigation keys to select Supervisor>Service>Diagnostic>SpO2>Error Counts.
3.
Select Reset CRC Counter. Select Yes when prompted to confirm the reset.
See “Functional Verification Tests” on page 1-22 to perform the functional verification tests.