C

HAPTER

2

T

ROUBLESHOOTING

34

9650-0804-01 Rev. A

QUICK ACCESS SW 1
SHORT

Quick access key 1 is stuck.

Replace the front panel.

QUICK ACCESS SW 2
SHORT

Quick access key 2 is stuck.

Replace the front panel.

QUICK ACCESS SW 3
SHORT

Quick access key 3 is stuck.

Replace the front panel.

QUICK ACCESS SW 4
SHORT

Quick access key 4 is stuck.

Replace the front panel.

QUICK ACCESS SW 5
SHORT

Quick access key 5 is stuck.

Replace the front panel.

QUICK ACCESS SW 6
SHORT

Quick access key 6 is stuck.

Replace the front panel.

QUICK ACCESS SW 7
SHORT

Quick access key 7 is stuck.

Replace the front panel.

SCREEN CYCLE SHORT

Display/home key is stuck.

Replace the front panel.

ALARM SILENCE SHORT

Alarm Silence/Reset key is
stuck.

Replace the front panel.

NAV PAD UP SHORT

Up/Clockwise Navigation
key is stuck.

Replace the front panel.

NAV PAD SELECT SHORT

Select key is stuck.

Replace the front panel.

NAV PAD DOWN SHORT

Down/Counterclockwise
Navigation key is stuck.

Replace the front panel.

SNAPSHOT SHORT

Snapshot key is stuck.

Replace the front panel.

ECG DEVICE FAILURE

ECG software detected a
problem with the ECG data
that is most likely a hardware
issue.

Replace the main board.

TEMP DEVICE FAILURE

Temp software detected a
problem with the
temperature data that is
most likely a hardware issue.

Replace the main board.

TEMP PROBE1 SHORTED

An A/D value was detected
that indicated the probe was
shorted.

Replace the temperature
probe or adapter cable. If
problem persists, replace
the right side panel.

Error Message

Explanation

Technical Action