Functional Verification Tests
9650-0804-01 Rev. A
ZOLL Propaq M Service Manual
29
IBP Test
Tools Needed
IBP patient simulator (with accuracy better than +/- 1 mmHg), or equivalent
Setup
Connect the test cables to IBP patient simulator.
Do This...
Observe This...
Pass/Fail
1.
Connect the IBP patient simulator to the
Propaq P1 channel.
2.
Set the IBP patient simulator to 0 mmHg.
3.
On the Propaq M, press > IBP.
4.
On the Propaq M, Press > P1->0<-.
Display of 0/0(+/-2) in the P1 box
at the bottom of the display (P1 is
zeroed).
The waveform for channel 1 is
present.
o
o
o
o
5.
Set the IBP patient simulator to 200
mmHg (or as close to 200 mmHg as your
simulator allows).
Display of 200/200(+/-5) in the P1
box at the bottom of the display as
xxx/yyy where yyy=diastolic (P1
diastolic pressure is 200 (+/-5)).
o
o
6.
Repeat steps 1 through 5 of this procedure
for P2 and P3 channels.