Functional Verification Tests

9650-0804-01 Rev. A

ZOLL Propaq M Service Manual

25

Heart Rate Alarm Test

Tools Needed

ECG simulator

Do this...

Observe this...

Pass/Fail

1.

Press the power button on the
Propaq M to turn the unit on.

Connect the ECG leads to the

ECG

simulator

. Set the simulator to 120

BPM and press the Leads quick

access key

(

)

on the Propaq M

unit and select lead II.

II displays next to the Leads quick
access key.

o

o

2.

On the Propaq M:

Press the Next quick

access key (

) key to

access the second set of
quick access keys.

Press the Alarm quick

access key (

).

Press the Limits softkey.

The Alarm Settings menu displays.

o

o

3.

On the Propaq M, select HR/PR
Alarm in the Alarm Settings menu
and use the navigation keys to do
the following:

Set the Upper Limit Enable

to ON.

Set the Lower Limit

Enable to ON.

Set the Lower limit to 30.

Set the Upper limit to 150.

Use the navigation keys to exit the
Alarm Settings menu.

4.

On the Propaq M, press the Alarm

Suspend key (

).

The red alarm suspension timer
displays at the top of the next to the
message area.

5.

Remove a lead wire from the

ECG

simulator

.

T

he ECG LEAD FAULT message

displays.

o

o

6.

On the Propaq M, press the Alarm

Suspend key (

).

The Alarm Suspend key flashes

.

7.

Reattach ECG Lead wire to

ECG

simulator

.

The ECG LEAD FAULT message no
longer displays.

o

o

Set simulator to 20 BPM or lower.

Heart Rate Value is highlighted and
the alarm tone sounds. The HR/PR
LOW ALARM
message displays.

8.

On the Propaq M, press the Alarm

Suspend key (

).

The Alarm Suspend key flashes.

The

red alarm suspension timer displays
at the top of the screen next to the
message area.

o

o

9.

Set the simulator to 120 BPM.

I, II,

III...