ZOLL Propaq M Service Manual

57

Chapter 4

Life Cycle Counters

Overview

This chapter contains instructions for setting life cycle counters on the Propaq M. The Propaq
M tracks various life cycle counters for specific modules installed in the device. When
replacing modules in the unit, it is important to reset the life cycle counters. Once you have
reset the appropriate life cycle counter(s), see “Functional Verification Tests” on page 1-22 to
perform the functional verification tests to ensure that the functions of the Propaq M unit work
properly.

Note:

Only reset the life cycle counters for the modules that you have replaced.

Resetting Life Cycle Counters

NIBP

On the Propaq M, do the following:

1. Press the Setup quick access key (

).

2.

Use the navigation keys to select Supervisor>Service>Diagnostic>NIBP>Cycle/Error
Counts
.

3.

Select Reset Counters. Select Yes when prompted to confirm the reset.

See “Functional Verification Tests” on page 1-22 to perform the functional verification tests.