Temperature Test

Tools Needed

IBP Temp Simulator, or equivalent

Test Setup

None

Do this...

Observe This...

Pass/Fail

1.

Connect the temperature simulator to the
Propaq M T1 channel.

2.

On the IBP Temp Simulator, Select 98.6
Degree F (36 Degree C) or a setting as close
as possible as your simulator allows.

The displayed temperature value is
within 0.2 F or 0.1 C of the
simulator setting.

o

o

3.

Repeat steps 1 and 2 of this procedure for the
T2 channel.