Propaq M Error Messages

9650-0804-01 Rev. A

ZOLL Propaq M Service Manual

37

NIBP AIR LEAK

Major air leak or could not
reach target pressure.

Attempt to identify
problems with the hose
and/or cuff by doing the
following:

Check the hose
connection to the
device. Make sure that
the hose is tightly
screwed onto the NIBP
connector.

Ensure that the cuff is
appropriately wrapped
and is not too loose.

Inspect the cuff and
hose for leaks. Inspect
the O-ring on the NIBP
hose connector.
Consider replacing the
cuff and/or hose.

If the problem persists,
perform one of the following
procedures in Chapter 1,
"NIBP Valve, Pump, and
Tubing Integrity Tests"
:

NIBP Leak Test

NIBP Inflation Test

Inspect internal NIBP
tubing for leaks.

NIBP EQUIPMENT CHECK

Device is unable to obtain a
reading.

Check the hose and cuff
connections. If problem
persists, perform one or
more of the following
procedures outlined in
Chapter 1, "NIBP Valve,
Pump, and Tubing Integrity
Tests"
:

NIBP Leak Test

NIBP Deflation Test

NIBP Inflation Test

NIBP WEAK PULSE

Too few pulses are detected
during the reading attempt.

Tighten the cuff and ensure
appropriate placement.

NIBP TEMP OUT OF
RANGE

Module temperature is out of
its specified operating range.

Move the device to a cooler
or warmer environment.

NIBP ARTIFACT

Unable to determine blood
pressure. Check
connections; restrict patient
movement.

Attempt to identify and
eliminate the source of
artifact. If problem persists,
replace the NIBP module.

Error Message

Explanation

Technical Action