Nibp test – ZOLL Propaq M Rev A User Manual
Page 34
C
HAPTER
1
M
AINTENANCE
T
ESTS
30
9650-0804-01 Rev. A
NIBP Test
Tools Needed
NIBP simulator, single or dual lumen hose, adult, NIBP cuff, rigid cylinder
Test Setup
Connect the Propaq M to the NIBP simulator using the single or dual lumen hose. Wrap the
NIBP cuff around a rigid cylinder, such as a glass jar.
Note:
Refer to Appendix B for information on simulators and settings that you may use to
assess the performance of the NIBP module.
Do This...
Observe This...
Pass/Fail
1.
On the Propaq M, ensure the patient mode
is set to Adult.
2.
Set the simulator to the 80/50 simulator
preset.
3.
Take a manual reading.
Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.
o
o
4.
Set the simulator to the 120/80 simulator
preset.
5.
Take a manual reading.
Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.
o
o
6.
Set the simulator to the 200/150 simulator
preset.
7.
Take a manual reading.
Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.
o
o