Nibp test – ZOLL Propaq M Rev A User Manual

Page 34

Advertising
background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

30

9650-0804-01 Rev. A

NIBP Test

Tools Needed

NIBP simulator, single or dual lumen hose, adult, NIBP cuff, rigid cylinder

Test Setup

Connect the Propaq M to the NIBP simulator using the single or dual lumen hose. Wrap the
NIBP cuff around a rigid cylinder, such as a glass jar.

Note:

Refer to Appendix B for information on simulators and settings that you may use to
assess the performance of the NIBP module.

Do This...

Observe This...

Pass/Fail

1.

On the Propaq M, ensure the patient mode
is set to Adult.

2.

Set the simulator to the 80/50 simulator
preset.

3.

Take a manual reading.

Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.

o

o

4.

Set the simulator to the 120/80 simulator
preset.

5.

Take a manual reading.

Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.

o

o

6.

Set the simulator to the 200/150 simulator
preset.

7.

Take a manual reading.

Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.

o

o