9650-0804-01 Rev. A

Heart Rate Display Test

Tools Needed

ECG simulator

ECG Cable (3-lead, 5-lead or 12-lead)

Test Setup

Press the power switch to turn on the Propaq M. Press Lead quick access key until II displays.

Connect the ECG leads to the ECG simulator.

Connect the ECG cable to the unit.

Do this...

Observe this...

Pass/Fail

1.

Set the ECG Simulator to 120BPM.

The Heart Rate displays as 120 +/- 4 bpm