Leads test – ZOLL Propaq M Rev A User Manual

Page 27

Advertising
background image

Functional Verification Tests

9650-0804-01 Rev. A

ZOLL Propaq M Service Manual

23

Leads Test

Tools Needed

3-lead, 5-lead, and 12-lead cables

Test each cable separately.

Test Setup

Connect the lead wires appropriate to each test to the

Fluke Impulse 7000DP Analyzer

or

equivalent.

Do this...

Observe this...

Pass/Fail

1.

Press the power switch to turn on the
Propaq M.

No ECG LEAD FAULT message
displayed.

o

o

2.

Disconnect one lead from the
simulator.

The ECG LEAD FAULT message displays
within 3 seconds (if configured).

o

o

3.

Reconnect the lead. Repeat step 2
with the remaining leads.

Wait for ECG LEAD FAULT message to
clear from the display (if configured).

o

o

4.

Repeat steps 2 and 3 for 5 lead and 12
lead cables.

Note: If heart rate alarm sounds, on the
Propaq M:

Press the Next quick access key

(

) key to access the second

set of quick access keys.

Press the Alarm quick access key

(

).

Press the Alarm Suspend key

(

).

o

o

Advertising