Leads test – ZOLL Propaq M Rev A User Manual
Page 27
Functional Verification Tests
9650-0804-01 Rev. A
ZOLL Propaq M Service Manual
23
Leads Test
Tools Needed
3-lead, 5-lead, and 12-lead cables
Test each cable separately.
Test Setup
Connect the lead wires appropriate to each test to the
Fluke Impulse 7000DP Analyzer
or
equivalent.
Do this...
Observe this...
Pass/Fail
1.
Press the power switch to turn on the
Propaq M.
No ECG LEAD FAULT message
displayed.
o
o
2.
Disconnect one lead from the
simulator.
The ECG LEAD FAULT message displays
within 3 seconds (if configured).
o
o
3.
Reconnect the lead. Repeat step 2
with the remaining leads.
Wait for ECG LEAD FAULT message to
clear from the display (if configured).
o
o
4.
Repeat steps 2 and 3 for 5 lead and 12
lead cables.
Note: If heart rate alarm sounds, on the
Propaq M:
•
Press the Next quick access key
(
) key to access the second
set of quick access keys.
•
Press the Alarm quick access key
(
).
•
Press the Alarm Suspend key
(
).
o
o