Dni nevada cufflink, Clinical dynamics smart arm – ZOLL Propaq M Rev A User Manual

Page 77

Advertising
background image

Overview

9650-0804-01 Rev. A

ZOLL Propaq M Service Manual

73

DNI Nevada CuffLink

Adult/Pediatric Mode

Clinical Dynamics Smart Arm

Adult/Pediatric Mode

Neonate Mode

FastBP

StepBP

Simulator
Setting

Expected
Systolic

Expected
Diastolic

Expected
Systolic

Expected
Diastolic

200/150

210

152

201

138

120/80

116

77

124

80

80/50

87

49

90

50

FastBP

StepBP

Simulator
Setting

Expected
Systolic

Expected
Diastolic

Expected
Systolic

Expected
Diastolic

200/150

192

152

194

136

120/80

124

79

122

80

80/50

85

42

80

49

NeoBP

Simulator
Setting

Expected
Systolic

Expected
Diastolic

80/50

77

48

60/30

57

32

35/15

31

13

Advertising