Dni nevada cufflink, Clinical dynamics smart arm – ZOLL Propaq M Rev A User Manual
Page 77
Advertising
Overview
9650-0804-01 Rev. A
ZOLL Propaq M Service Manual
73
DNI Nevada CuffLink
Adult/Pediatric Mode
Clinical Dynamics Smart Arm
Adult/Pediatric Mode
Neonate Mode
FastBP
StepBP
Simulator
Setting
Expected
Systolic
Expected
Diastolic
Expected
Systolic
Expected
Diastolic
200/150
210
152
201
138
120/80
116
77
124
80
80/50
87
49
90
50
FastBP
StepBP
Simulator
Setting
Expected
Systolic
Expected
Diastolic
Expected
Systolic
Expected
Diastolic
200/150
192
152
194
136
120/80
124
79
122
80
80/50
85
42
80
49
NeoBP
Simulator
Setting
Expected
Systolic
Expected
Diastolic
80/50
77
48
60/30
57
32
35/15
31
13