Functional Verification Tests

Before You Begin the Functional Verification Tests

•

Assemble the tools listed in following section, “Equipment You Need to Perform the

Functional Verification Tests.”

•

Keep an extra fully charged Propaq M battery available.

•

Schedule an hour to conduct the entire maintenance test.

•

Perform the tests in the order presented.

•

Perform all the steps of each test procedure.

•

Complete all the steps of the procedure before evaluating the test results.

Equipment You Need to Perform the Functional Verification Tests

For testing purposes, you can substitute an equivalent device.

•

Fluke Impulse 7000DP Analyzer, or equivalent, with appropriate interface adapter

(Dynatech Nevada Inc. part number 3010-0378)

•

IBP/Temp Simulator, or equivalent

•

NIBP Simulator

•

Single or dual lumen NIBP hose

•

Adult NIBP cuff and a rigid cylinder to wrap it around

•

5% CO

2

calibration gas cylinder with CO

2

sample line (kit available from Scott Medical

Products, PN 4653ORF-BD)

•

ECG Simulator; 12 Lead Simulator for 12 Lead test (e.g., Symbio CS1201)

Accessories You Need for the Propaq M Options Functional Verification
Tests

•

3-lead, 5-lead and 12-lead ECG cables. (12-lead cable needed if 12-lead option is

installed.)

•

NIBP cuff and hose

•

IBP cable

•

SpO

2

cable

and sensor

•

EtCO

2

cable

and sensor

•

Paddles

•

Battery

•

Auxiliary power source