ZOLL Propaq M Rev A User Manual

Page 56

Advertising
background image

52

9650-0804-01 Rev. A

3

Disconnect multi-wire IBP cable with ferrite
in the area illustrated.

4

Cut the tie wrap to the IBP cable in the area
shown with the orange stick.

Note: During reassembly, secure a new tie
wrap to the IBP cable.

5

Disconnect the USB flex cable from ZIF
connector in the area illustrated.

6

Lay front end of separated unit down with
opened side facing up.

7

Disconnect the black EtCO

2

tube

connector from the air vent.

8

Remove four (4) screws in the areas
illustrated to release the silver EtCO

2

module.

Note: During reassembly, torque screws to
6 in-lbs.

multi-wire IBP cable

Advertising