ZOLL Propaq M Rev A User Manual
Page 43
Advertising
Propaq M Error Messages
9650-0804-01 Rev. A
ZOLL Propaq M Service Manual
39
CO2 GAS INPUT
OCCLUSION
CO
2
inlet blocked.
Replace FilterLine
®
. If
problem persists, check for
CO
2
inlet occlusion. If still
persists, replace the CO
2
module.
CO2 MAINTENANCE
REMINDER
Periodic maintenance
reminder.
Replace CO
2
module soon.
CO2 MAINTENANCE
REQUIRED
Periodic Maintenance due.
Replace CO
2
module.
CO2 NO FILTERLINE
No filterline is detected.
If FilterLine is connected
and problem persists,
replace FilterLine. If still
persists, replace the CO
2
module.
CO2 CAL FAILURE
Calibration failed.
Retry calibration or
continue with old
calibration. Replace the
CO
2
module.
Error Message
Explanation
Technical Action