ZOLL Propaq M Rev A User Manual

Page 43

Advertising
background image

Propaq M Error Messages

9650-0804-01 Rev. A

ZOLL Propaq M Service Manual

39

CO2 GAS INPUT
OCCLUSION

CO

2

inlet blocked.

Replace FilterLine

®

. If

problem persists, check for
CO

2

inlet occlusion. If still

persists, replace the CO

2

module.

CO2 MAINTENANCE
REMINDER

Periodic maintenance
reminder.

Replace CO

2

module soon.

CO2 MAINTENANCE
REQUIRED

Periodic Maintenance due.

Replace CO

2

module.

CO2 NO FILTERLINE

No filterline is detected.

If FilterLine is connected
and problem persists,
replace FilterLine. If still
persists, replace the CO

2

module.

CO2 CAL FAILURE

Calibration failed.

Retry calibration or
continue with old
calibration. Replace the
CO

2

module.

Error Message

Explanation

Technical Action

Advertising