<4. Diagnostics>

4-2

IM 01C25R02-01E

4.1.2 Checking with Integral Indicator

NOTE

If an error is detected by running self-diagnostics,

an error number is displayed on the integral

indicator. If there is more than one error, the error

number changes at three-second intervals.

See table 4.3.1 regarding the alarm codes.

F0401.ai

Figure 4.1

Integral Indicator

4.1.3 Status information available for

HART 7

Status added to HART 7 is explained below.

(1) Device Status

Device Status indicates the current operating

status of the device. (Refer to Table 4.7)

Table 4.10 indicates the relationship between

alarm and Device Status.

• Procedure to call up the display using DD

(HART 7) and DTM (HART 7)

DD (HART 7)

DTM (HART 7)

[Root Menu] → Diag/Service →

Status → Device Status

(2) Extended Device Status

Ext dev status contains commonly used

device information. (Refer to Table 4.8)

Table 4.10 indicates the relationship between

alarm and Extended Device Status.

• Procedure to call up the display using DD

(HART 7) and DTM (HART 7)

DD (HART 7)

DTM (HART 7)

[Root Menu] → Diag/Service →

Status → Ext dev status

(3) Data quality and Limit status

EJX multivariable transmitter can handle

Dynamic Variables (PV, SV, TV, QV) and Device

Variables (DP(Pres), SP, ET, Flow, Total Flow,

% range, Loop current). Each variable contains

data quality and limit status for providing useful

status about the data value. The data quality

is normally “Good”. However, in the case of a

sensor failure or out of measurement range,

it turns to “Bad” or “Poor Accuracy”. The limit

status indicates whether the data value is

limited (i.e., not responding to the process).

When the limit status is “Constant”, the value

will not be changed. For detail, refer to Table 4.9

and 4.10.

• Procedure to call up the display using DD

(HART 7) and DTM (HART 7)

[Dynamic Variables]

DD (HART 7)

DTM (HART 7)

[Root Menu] → Process variables

→ Output vars → PV →

→ PV Data

Quality

Good, Poor Accuracy, Manual/

Fixed, or Bad is displayed.

→ PV Limit

Status

Constant, Low Limit, High Limit, or

Not Limited is displayed.

It is the same about the SV, TV, and QV.

[Device Variables]

DD (HART 7)

DTM (HART 7)

[Root Menu] → Process variables

→ Device variables and Status →

→ Flow Data

Quality

Good, Poor Accuracy, Manual/

Fixed, or Bad is displayed.

→ Flow Limit

Status

Constant, Low Limit, High Limit, or

Not Limited is displayed.

It is the same about the Pres, SP, ET, Total Flow, %

Range, and Loop Current. .

(4) Time Stamp

Time Stamp displays the date and the time

information which EJX multivariable transmitter

maintains from the time of the power on.

It is used as the additional information of the

process value and the event.

• Procedure to call up the display using DD

(HART 7) and DTM (HART 7)

DD (HART 7)

DTM (HART 7)

[Root Menu] → Diag/Service →

Status → Time Stamp

→ Current Date

It shows the number of operating

days.

→ Current Time

It shows the running time.