<4. Diagnostics>

4-4

IM 01C25R02-01E

IMPORTANT

• The pressure fluctuation amplitude in fluids

must be sufficiently large for blockages to be

detected.

• If the pressure fluctuation amplitude is too

low for a reference value to be obtained,

blockages detection operation cannot be

performed with an alarm that the reference

value is invalid.

• The pressure fluctuation amplitude may

decrease due to other causes unrelated with

a blockage according to process condition. In

above case, a false alarm of an impulse line

blockage may be generated. Before taking

action in response to a blockage alarm,

consider the plant operating conditions.

Notes for Pressure or Level

Measurement

With pressure or level measurement, the pressure

fluctuation amplitude may reduce especially for the

following cases.

Pressure Measurement

• Operational pressure is near outside of

diagnostic range.

• Even though pressure is constant, the flow

decreases than that under normal condition.

• A source of pressure fluctuation (pump,

compressor, blower, etc.) is shut down. As

a result, the pressure fluctuation amplitude

decreases.

Level Measurement

• A transmitter is used to measure tank level and

the flow of fluid into or out of the tank comes to

a stop.

• The agitator in the tank is shut down.

• A source of pressure variation (a compressor,

etc.) that controls the internal pressure of a

sealed (closed) tank is shut down.

Before taking action in response to a blockage

alarm, consider the plant operating conditions.