Yokogawa EJX930A User Manual

Page 78

Advertising
background image

<4. Diagnostics>

4-28

IM 01C25R02-01E

Table 4.10

Relationship between Alarm and Status available for HART 7

Integral

Indicator

HART

configuration

tool display

Device

Status

Extended

Device Status

Value and Status (Data Quality and Limit Status)

Flow

Differential

Pressure

(DP)

Static

Pressure

(SP)

External

Temperature

(ET)

Total Flow

% range

Loop

current

AL.01

CAP.ERR

P sensor

error

Device

Malfunction

(0x80)

Maintenance

Required

(0x01)

Value: Hold value

Status: Bad and Constant

Value:Hold value

Status:Bad and Low

Limited/High Limited

*1

CT sensor

error

Value: Hold value

Status: Bad and Constant

Cap

EEPROM

error

Value: Hold value

Status: Bad and Constant

AL.02

AMP.ERR

AT sensor

error

Value: Hold value

Status: Bad and Constant

Value: Hold value

Status: Bad and Low

Limited/High Limited

*1

Amp

EEPROM

error

Value: Hold value

Status: Bad and Constant

CPU board

error

Value: Hold value

Status: Bad and Constant

AD

Converter

error

Value: Hold value

Status: Bad and Constant

AL.03

ET. ERR

ET sensor

error

Value: Hold value

Status: Bad and Constant

Value: Hold value

Status: Bad and Low

Limited/High Limited

*1

---

No device

ID

---

---

Value: Measured value

Status: Good and Not Limited

AL.10

PRESS

P outside

limit

In the case

of PV:

Primary

Variable

Out of

Limits

(0x01)

In the case

except PV:

Non-

Primary

Variable

Out of

Limits

(0x02)

Device

Variable Alert

(0x02)

Value:

Measured

value

Status:

Poor

Accuracy

and

Not Limited

Value:

Measured value

Status:

Poor Accuracy and Not

Limited

Value:

Measured

value

Status:

Good and

Not Limited

Value:

Measured

value

Status:

Poor

Accuracy

and

Not Limited

Same as Value and

Status of PV

AL.11

ST.PRSS

SP outside

limit

AL.12

CAP.TMP

CT outside

limit

Non-

Primary

Variable

Out of

Limits

(0x02)

AL.13

AMP.TMP

AT outside

limit

Value: Measured value

Status: Good and Not Limited

AL.14

EXT.TMP

ET outside

limit

In the case

of PV:

Primary

Variable

Out of

Limits

(0x01)

In the case

except PV:

Non-

Primary

Variable

Out of

Limits

(0x02)

Value:

Measured

value

Status:

Poor

Accuracy

and

Not Limited

Value: Measured value

Status: Good and Not

Limited

Value:

Measured

value

Status: Poor

Accuracy

and Not

Limited

Value:

Measured

value

Status:

Poor

Accuracy

and

Not Limited

Same as Value and

Status of PV

AL.15

EXT.TMP

OHM

outside limit

AL.16

PLS

PLS outside

limit

---

Value: Measured value

Status: Good and Not Limited

*1: Depend on the setting of hardware switch

*2: Depend on the direction of range over (high or low)

Advertising