Yokogawa EJX930A User Manual

Page 79

Advertising
background image

<4. Diagnostics>

4-29

IM 01C25R02-01E

Integral

Indicator

HART

configuration

tool display

Device

Status

Extended

Device Status

Value and Status (Data Quality and Limit Status)

Flow

Differential

Pressure

(DP)

Static

Pressure

(SP)

External

Temperature

(ET)

Total Flow

% range

Loop

current

AL.30

PRS.RNG

P over

range

In the case

of PV:

Loop

Current

Saturated

(0x04)

Device

Variable Alert

(0x02)

Value: Measured value

Status: Good and Not Limited

In the case of PV

Value: Hold value

Status: Bad and

Low Limited/High

Limited

*2

In the case except PV

Value: Measured

value

Status: Good and

Not Limited

AL.31

SP. RNG

SP over

range

AL.32

F. RNG

F over

range

AL.33

ET. RNG

ET over

range

AL.35

P. HI

P high

alarm

---

Value: Measured value

Status: Good and Not Limited

AL.36

P. LO

P low alarm

AL.37

SP. HI

SP high

alarm

AL.38

SP. LO

SP low

alarm

AL.41

F. HI

F high

alarm

AL.42

F. LO

F low alarm

AL.43

ET. HI

ET high

alarm

AL.44

ET. LO

ET low

alarm

AL.50

P. LRV

Illegal P

LRV

Value: Measured value

Status: Good and Not Limited

In the case PV is DP

Value: Hold value

Status: Bad and

Constant

In the case except

shown above

Value: Measured

value

Status: Good and

Not Limited

AL.51

P. URV

Illegal P

URV

AL.52

P. SPN

Illegal P

SPAN

AL.53

P. ADJ

P SPAN

trim err

Value:

Measured

value

Status:

Poor

Accuracy

and

Not Limited

Value:

Measured

value

Status:

Poor

Accuracy

and

Not Limited

Value:

Measured value

Status:

Good and Not Limited

Value:

Measured

value

Status:

Poor

Accuracy

and

Not Limited

In the case PV is DP

or Flow

Value: Measured

value

Status: Poor

Accuracy and Not

Limited

In the case except

shown above

Value: Measured

value

Status: Good and

Not Limited

P ZERO

trim err

AL.54

SP. RNG

Illegal SP

LRV

Value: Measured value

Status: Good and Not Limited

In the case PV is SP

Value: Hold value

Status: Bad and

Constant

In the case except

shown above

Value: Measured

value

Status: Good and

Not Limited

Illegal SP

URV

Illegal SP

SPAN

*1: Depend on the setting of hardware switch

*2: Depend on the direction of range over (high or low)