2 combination of reference result, And blockage detection -8 – Yokogawa EJX930A User Manual

Page 58

Advertising
background image

<4. Diagnostics>

4-8

IM 01C25R02-01E

NOTE

Ref fDP must be larger than the specified

level shown in Table 4.4 (refer to subsection

4.2.2.6). No blockage can be detected when

Ref fDP is not large enough.

• The plausibility of blockage detection needs

to be confirmed by blockage simulation test.

The simulation test can be performed by

the appropriate manifold operation (refer to

subsection 4.2.2.8).

4.2.2.2 Combination of Reference Result

and Blockage Detection

Diag Applicable

An EJX multivariable transmitter can detect four

modes of impulse line blockage: both-sides, high-

pressure side, low-pressure side, and/or single-side

and abnormal fluctuation when all the reference

values are properly measured. However, the

detectable alarm mode combination is limited when

some of the reference values are invalid. Available

Blockage Detection are shown in the below figure.

Ref fDP:

OK

Ref fSPI:

OK

Ref fSPh:

OK

Ref BlkF:

OK

Fluctuation Parameters

• Test for low-pressure side

• Test for high-pressure side

• Test for both-pressure sides

Simulation Test

• B Blocking detection

(by using

Ratio fDP)

• L Side Blocking detection

(by using

Ratio fSPI)

• H Side Blocking detection

(by using

Ratio fSPh)

• H/L Side Blocking detection

(by using

BlkF)

Ref fDP:

OK

Ref fSPI:

OK

Ref fSPh:

OK

Ref BlkF:

NG

• Test for low-pressure side

• Test for high-pressure side

• Test for both-pressure sides

• A/B Blocking detection

(by using

Ratio fDP)

• L Side Blocking detection

(by using

Ratio fSPl)

• H Side Blocking detection

(by using

Ratio fSPh)

Ref fDP:

OK

Ref fSPI:

NG

Ref fSPh:

OK

Ref BlkF:

NG

• Test for high-pressure side

• A Blocking detection

(by using

Ratio fDP)

• H Side Blocking detection

(by using

Ratio fSPh)

Ref fDP:

OK

Ref fSPI:

NG

Ref fSPh:

NG

Ref BlkF:

NG

• Test for both-pressure sides

• B Blocking detection

(by using

Ratio fDP)

Available Blockage Detection

[Differential pressure measurement]

[Gauge/absolute pressure and level measurement]

Ref fDP:

OK

Fluctuation Parameters

Simulation Test

• Test for high-pressure side

• B Blocking detection

(by using

Ratio fDP)

Available Blockage Detection

F0405.ai

Advertising
This manual is related to the following products: