Yokogawa EJX930A User Manual

Page 81

Advertising
background image

<4. Diagnostics>

4-31

IM 01C25R02-01E

Integral

Indicator

HART

configuration

tool display

Device

Status

Extended

Device Status

Value and Status (Data Quality and Limit Status)

Flow

Differential

Pressure

(DP)

Static

Pressure

(SP)

External

Temperature

(ET)

Total Flow

% range

Loop

current

AL.87

FLG. HI

FT high

alarm

---

Maintenance

Required

(0x01)

Value: Measured value

Status: Good and Not Limited

In the case “Off” is set

to Diag Out Option

Value: Measured

value

Status: Good and

Not Limited

In the case “Burnout”

is set to Diag Out

Option

Value: Low Limited/

High Limited

Status: Bad and

Low Limited/High

Limited

*1

In the case “Fall back”

is set to Diag Out

Option

Value: Fixed value

Status: Bad and

Constant

AL.87

FLG. LO

FT low

alarm

AL.88

INVR.DP

Invalid Ref

DP

Value: Measured value

Status: Good and Not Limited

AL.88

INVR.SL

Invalid Ref

SPL

AL.88

INVR.SH

Invalid Ref

SPH

AL.88

INVR.F

Invalid

Ref F

AL.89

ILBD.OV

ILBD over

range

Value: Measured value

Status: Good and Not Limited

AL.89

B BLK

B Blocking

Value: Measured value

Status: Good and Not Limited

In the case “Off” is set

to Diag Out Option

Value: Measured

value

Status: Good and

Not Limited

In the case “Burnout”

is set to Diag Out

Option

Value: Low Limited/

High Limited

Status: Bad and

Low Limited/High

Limited

*1

In the case “Fall back”

is set to Diag Out

Option

Value: Fixed value

Status: Bad and

Constant

AL.89

H BLK

H Side

Blocking

AL.89

L BLK

L Side

Blocking

AL.89

H LRG

Large Fluct

H

AL.89

L LRG

Large Fluct

L

AL.89

A BLK

A Blocking

AL.90

SIM

Flow

Simulation

mode

---

Device

Variable Alert

(0x02)

Value:

Measured

value

Status:

Good and

Not Limited

Value: Measured value

Status: Good and Not Limited

or

Value: Simulation value

Status: Manual/Fixed and Constant

Value:

Measured

value

Status:

Good and

Not Limited

Same as Value and

Status of PV

*1: Depend on the setting of hardware switch

*2: Depend on the direction of range over (high or low)