<4. Diagnostics>

4-6

IM 01C25R02-01E

4.2.2.1 Blockage Detection

Limit parameter

When the parameter based on pressure fluctuation

exceeds the preset value, EJX diagnoses an

impulse line as blockage and gives an alarm. The

threshold values are set to Limit parameter shown

in below table.

• Procedure to call up the display

DD (HART 5/7)

DTM (HART 7)

[Root Menu] → Diag/Service → Diag

Parameters → ILBD Parameters →

Configuration → Diag Lim →

DTM (HART 5) Diag and Service → Advanced Diag

Configurations → Impulse Line

Blockage Detection → Threshold

→ Sensitivity →

Table 4.2

Limit Parameter

#

Parameter

Threshold value

[1] Lim fDPmax

Threshold to detect “A

Blocking” by using

Ratio fDP

[2] Lim fDPmin

Threshold to detect “B

Blocking” by using

Ratio fDP

[3] Lim fSPlmax

Threshold to detect “Large

Fluct L” by using

Ratio fSPl

[4] Lim fSPlmin

Threshold to detect “L Side

Blocking” by using

Ratio fSPl

[5] Lim fSPhmax

Threshold to detect “Large

Fluct H” by using

Ratio fSPh

[6] Lim fSPhmin

Threshold to detect “H Side

Blocking” by using

Ratio fSPh

[7] Lim BlkFmax

Threshold to detect “H Side

Blocking” by using

BlkF

[8] Lim BlkFmin

Threshold to detect “L Side

Blocking” by using

BlkF

[9] Lim DPAvgmax Threshold to detect “ILDB over

range” by using

DPAvg and to

detect “Invalid Ref xx” by using

Ref DPAvg

[10] Lim DPAvgmin

Threshold to detect “ILDB over

range” by using

DPAvg and to

detect “Invalid Ref xx” by using

Ref DPAvg

Table 4.3 shows the default values at the factory

setting.

NOTE

• When ILBD is performed for the first time,

use the default value. If the pressure

fluctuation amplitude is low or a false alarm

is often generated after ILBD is performed,

change the values of Limit parameters

according to the procedure described in

subsection 4.2.2.10. Tuning

Table 4.3

Default Values of Limit Parameter

#

Parameter

Multivariable Transmitter
EJX910A, EJX930A

[1] Lim fDPmax

3

[2]

Lim fDPmin

0.3

[3]

Lim fSPlmax

5

[4]

Lim fSPlmin

0.5

[5]

Lim fSPhmax

5

[6]

Lim fSPhmin

0.5

[7]

Lim BlkFmax

0.6

[8]

Lim BlkFmin

-0.6

[9]

Lim DPAvgmax

Note 1

1

[10] Lim DPAvgmin

Note 1

0.05

Note 1: It indicates the threshold value for “ILBD over range”

(refer to subsection 4.2.2.5).

A/B Blocking Detection

“A Blocking” and “B Blocking” indicates the result

estimated from blockage degree based on the

difference of the high- and low-pressure-side

fluctuation values.

Ratio fDP, SQRT (fDP / Ref

fDP) is used to detect A/B blocking. Ref fDP is the

average value of the sum of squares of differential

pressure fluctuations under normal condition.
As the value of

Ratio fDP exceeds the value of

Lim fDPmax, EJX gives basically an alarm of “A

Blocking”. On the other hand, if this value is below

the value of

Lim fDPmin, EJX gives an alarm of “B

Blocking.
As a high- or low-pressure-side blockage

progresses,

fDP increases. Therefore, “A Blocking”

with a differential pressure transmitter indicates

that a single-side impulse line is plugged for a

differential pressure transmitter. As the both-side

blockages progress simultaneously,

fDP decreases.

Therefore, “B Blocking” with a differential pressure

transmitter indicates that both-side impulse lines

are plugged.

NOTE

A single-side impulse line blockage may

generate “B blocking” under the condition where

the fluctuation amplitude is much different

between high- and low-pressure sides.