Yokogawa EJX930A User Manual

Page 74

Advertising
background image

<4. Diagnostics>

4-24

IM 01C25R02-01E

Integral

indicator

HART configuration

tool display

Cause

4-20mA Output

operation during error

Countermeasure

Status

group

*

1

Diagnostic

List

group *

2

AL. 30

PRS.RNG

P over range *

1

Pressure Over Range

*

2

Differential pressure

exceeds specified range.

When PV is Pres.

Output AO upper limit or

lower limit.

(Refer to subsection 3.3.1)

Check input and

range setting, and

change them as

needed.

4

Tr

an

sd

uc

er

S

ta

tu

s

AL. 31

SP. RNG

SP over range *

1

Static Pressure Over

Range *

2

Static pressure exceeds

specified range.

When PV is SP

Output AO upper limit or

lower limit.

(Refer to subsection 3.3.1)

AL. 32

F. RNG

F over range *

1

Flow Over Range *

2

Flow exceeds specified

range.

When PV is Flow

Output AO upper limit or

lower limit.

(Refer to subsection 3.3.1)

AL. 33

ET. RNG

ET over range *

1

External Temp Over

Range *

2

External temperature

exceeds specified range.

When PV is ET

Output AO upper limit or

lower limit.

(Refer to subsection 3.3.1)

AL. 35 P.HI P high alarm *

1

Pressure High Alarm *

2

Input pressure exceeds

specified threshold.

Continues to operate and

output.

Check input.

5

AL. 36 P.LO P low alarm *

1

Pressure Low Alarm *

2

AL. 37

SP.HI

SP high alarm *

1

Static Pressure High

Alarm *

2

Input static pressure

exceeds specified

threshold.

AL. 38

SP.LO

SP low alarm *

1

Static Pressure Low

Alarm *

2

AL. 41 F.HI F high alarm *

1

Flow High Alarm *

2

Input flow exceeds

specified threshold.

8

AL. 42 F.LO F low alarm *

1

Flow Low Alarm *

2

AL. 43

ET.HI

ET high alarm *

1

External Temp High

Alarm *

2

Input external

temperature exceeds

specified threshold.

AL. 44

ET.LO

ET low alarm *

1

External Temp Low

Alarm *

2

AL. 50

P. LRV

Illegal P LRV *

1

Illegal Pressure LRV

*

2

Specified value is outside

of setting range.

Holds at the output value

that existed immediately

before the error occurred.

Check settings and

change them as

needed.

6

C

on

fig

ur

at

io

n

AL. 51

P. URV

Illegal P URV *

1

Illegal Pressure URV

*

2

AL. 52

P. SPN

Illegal P SPAN *

1

Illegal Pressure SPAN

*

2

AL. 53

P. ADJ

P SPAN trim err *

1

Pressure SPAN Trim

Error *

2

Continues to operate and

output.

Adjust settings and

change them as

needed.

P ZERO trim err *

1

Pressure ZERO Trim

Error *

2

AL. 54

SP. RNG

Illegal SP LRV *

1

Illegal Static Pressure

LRV *

2

Holds at the output value

that existed immediately

before the error occurred.

Check settings and

change them as

needed.

Illegal SP URV *

1

Illegal Static Pressure

URV *

2

Illegal SP SPAN *

1

Illegal Static Pressure

SPAN *

2

*1: DD (HART 5/HART 7), DTM (HART 7)

*2: DTM (HART 5)

*3: Available only for HART 7

Advertising