A1.3 Transducer Block

relative

Index

Index

Parameter Name

Factory default

Write

Mode

Explanation

0

2000

Block Header

tag: tB

Block
tag=o/S

Information on this block such as Block tag,
DD Revision, Execution time etc.

1

2001

St_REV

-

the revision level of the static data associated
with the function block. the revision value
will be incremented each time a static
parameter value in the block is changed.

2

2002

taG_DESC

(Spaces)

auto

the user description of the intended appli-
cation of the block

3

2003

StRatEGY

1

auto

the strategy field can be used to identify
grouping of blocks. this data is not checked
or processed by the block.

4

2004

aLERt_KEY

1

auto

the identification number of the plant unit.
this information may be used in the host for
sorting alarms, etc.

5

2005

MoDE_BLK

auto

auto

the actual, target, permitted, and normal
modes of the block.

6

2006

BLoCK_ERR

0

-

this parameter reflects the error status as-
sociated with hardware or software
components associated with a block. It is
a bit string, so that multiple errors may be
shown.

7

2007

uPDatE_EVt

-

this alert is generated by any change to the
static data

8

2008

BLoCK_aLM

-

the block alarm is used for all configura-
tion, hardware, connection failure or system
problems in the block. the cause of the
alert is entered in the sub-code field. the
first alert to become active will set the active
status in the Status attribute.

9

2009

tRanSDuCER_DIRECtoRY

1,2010

-

a directory that specifies the number and
starting indices of the device.

10

2010

tRanSDuCER_tYPE

Standard flow
with calibration
(104)

-

Identifies the device type, which is “Stand-
ard Flow with Calibration” for the Rotamass.

11

2011

XD_ERRoR

0 (no Error)

-

Indicates the error code of the error of the
highest priority from among the errors
currently occurring in the transducer block.
0=no failure, Range 127 (CPu-failure) to
100 (autozero out of Range)

12

2012

CoLLECtIon_DIRECtoRY

-

a directory that specifies the number, start-
ing indices, and DD Item IDS of the data
collections in each transducer within a
transducer block.

13

2013

CaL_PoInt_HI

*)

o/S

the highest calibrated value. to set within
the range of SEnSoR_RanGE.

14

2014

CaL_PoInt_Lo

0

o/S

the lowest calibrated value. to set within
the range of SEnSoR_RanGE.

15

2015

CaL_MIn_SPan

1500

-

the minimum calibration span value allowed.
10% of SEnSoR_RanGE.

16

2016

CaL_unIt

kg/h

-

the engineering unit for the calibrated
sensor.

17

2017

SEnSoR_tYPE

Coriolis (101)

-

Indicates the sensor type, which is “Coriolis”
for the Rotamass.

18

2018

SEnSoR_RanGE

*)

-

the high and low range limits values,
engineering units code and the number of
digits to the right of the decimal point for the
sensor.

19

2019

SEnSoR_Sn

Serial no.

-

the serial number of the connected sensor.

20

2020

SEnSoR_CaL_MEtHoD

Dynamic weigh
(102)

o/S

the method of the last sensor calibration

21

2021

SEnSoR_CaL_LoC

YoKoGaWa

o/S

Sets/indicates the location of the last sensor
calibration.