5. CONFIGUrATION

5-5

IM 01R04B05-00E-E 3rd edition July 30 2010 -00

all Rights Reserved. Copyright © 2005, Rota Yokogawa

T0504-2.EPS

13

FasDllSubscriber
SynchronizationDlcep

Not used for the Rotamass.

14

FmsVfdId

Sets VFD for the Rotamass
to be used.

0x1: System/network

management VFD

0x1234: Function block

VFD

15

FmsMaxOutstanding
ServiceCalling

Set 0 to Server. It is not
used for other applications.

16

FmsMaxOutstanding
ServiceCalled

Set 1 to Server. It is not
used for other applications.

17

FmsFeatures
Supported

Indicates the type of services
in the application layer. In the
Rotamass, it is automatically
set according to specific
applications.

S

ub

-

index

P

arameter

D

escription

12

FasDllSubscriberTime
WindowSize

Not used for the Rotamass.

these 33 VCRs are factory-set as shown in table
5.5.

Table 5.5 VCr List

T0505.EPS

I

ndex

(SM)

VCR

N

umber

F

actory

S

etting

293

For system management (Fixed)

1

294

Server (LocalAddr = 0xF3)

2

295

Server (LocalAddr = 0xF4)

3

296

Server (LocalAddr = 0xF7)

4

297

Trend Source (LocalAddr = 0x07,
Remote Address=0x111)

5

298

Publisher (LocalAddr = 0x20)

6

299

Alert Source (LocalAddr = 0x07,
Remote Address=0x110)

7

300

Server (LocalAddr = 0xF9)

8

9 to 33

Not set

301 to 325

5.5.2 Function Block Execution Control

according to the instructions given in Section 5.3,
set the execution cycle of the function blocks and
schedule of execution.

5.6 Block Setting

Set the parameter for function block VFD.

5.6.1 Link Objects

a link object combines the data voluntarily sent by
the function block with VCR. Each RotaMaSS
has 40 link objects. a single link object speci-
fies one combination. Each link object has the
parameters listed in table 5.6. Parameters must
be changed together for each VCR because the
modifications made to each parameter may cause
inconsistent operation.
Table 5.6 Link Object Parameters

T0506.EPS

S

ub

-

index

P

arameters

D

escription

1

LocalIndex

Sets the index of function
block parameters to be
combined; set “0” for Trend
and Alert.

2

VcrNumber

Sets the index of VCR to
be combined. If set to “0”,
this link object is not used.

3

RemoteIndex

Not used in the Rotamass.
Set to “0”.

5

StaleCountLimit

4

ServiceOperation

Set one of the following.
Set only one each for link
object for Alert or Trend.
0: Undefined
2: Publisher
3: Subscriber
6: Alert
7: Trend

Set the maximum number
of consecutive stale input
values which may be
received before the input
status is set to BAD. To
avoid the unnecessary
mode transition caused
when the data is not
correctly received by
subscriber, set this
parameter to “2” or more.

Link objects are not factory-set. Set link objects as
shown in table 5.7.
Table 5.7 Settings of Link Objects (example)

t0507.EPS

Index

Link object #

Settings(example)

30000

aI. out VCR#6

1

30001

2

30002

3

30003 to 30039

no used

4 to 40

trend VCR#5

alert VCR#7