8. GENErAL SPECIFICATIONS

8-8

IM 01R04B05-00E-E 3rd edition July 30, 2010 -00

all Rights Reserved. Copyright © 2005, Rota Yokogawa

FACTOry SETTING

Parameter legend:
(1) XD_SCaLE:

Defines the input values from the transducer block (input range of sensor) corresponding to 0% and

100% span values from the inside calculation of the aI function blocks.

the values set as the mass flow span, volume flow span, density span and temperature span are

stored in this parameter in the RCCt3/RCCF31.

Concentration span and net flow span can be set in this parameter under option /Cxx.

(2) out_SCaLE: output scaling parameter. Defines the output values corresponding to 0% and 100% span values

from the inside calculation of the aI function blocks.

(3) PV_FtIME: time constant of the damping function within the aI blocks is set to 0 s.
(4) L_tYPE:

Determines if the values passed by the transducer block to the aI block may be used directly (Direct)

or if the value is in different units and must be converted linearly (Indirect Linear) using the input range

defined by XD_SCaLE and the associated output range (out_SCaLE).

Item

Settings

tag number (tag plate, option /BG)

as specified in order

*1

Software tag (PD_taG)

Set to “Ft1004” by default unless otherwise specified when ordered

*2

node address

Set to 0xF6 (246) by default unless otherwise specified when ordered

*3


*1

Specified tag number is engraved on the stainless steel plate: up to 16 letters using any alphanumeric and symbols of {-},{.} and {/}.

*2

Specified software tag number is entered in the amplifier memory: up to 32 letters using any alphanumeric and symbols of {-},{.} and {/}.

*3

Range of node address: 0x00 to 0xFF (0 to 255).

Item

Settings

operation Functional Class

Set to ‘BaSIC´ unless otherwise specified when ordered

analog Input Function Block

aI1 Mass Flow

aI2 Volume Flow

aI3 Density

aI4 temperature

upper and lower operating range limits and unit
(XD_SCaLE)

the range limits will be
set to the mass flow rate
range specified on the
order sheet (/PS) or to
0 to Qmax

*1

if the order

sheet is not supplied.

the range limits will be
set to the volume flow
rate range specified on
the order sheet (/PS) or
to 0 to Qvmax

*2

if the

order sheet is not
supplied.

the range limits will be
set to the density range
specified on the order
sheet (/PS) or to 0 to 1.5
kg/l if the order sheet is
not supplied.

the range limits will
be set to the tempera-
ture range specified
on the order sheet
(/PS) or to 0 to 150 °C
if the order sheet is
not supplied.

upper and lower output range limits and unit
(out_SCaLE)

Damping time constant (tB-Block)

3 s

3 s

3 s

3 s

analog Input Function Block

aI5 Concentration Measurement

aI6 net Flow

upper and lower operating range limits and unit
(XD_SCaLE)

the range limits will be set to the concentration
measurement range specified on the order sheet
(/PS) or to 0 to 100 Wt-% if the order sheet is
not supplied. the unit depends on the selected
concentration.

the range limits will be set to the net flow rate
range specified on the order sheet (/PS) or to
0 to Qmax

*1

if the order sheet is not supplied.

upper and lower output range limits and unit
out_SCaLE)

Damping time constant (tB-Block)

10 s

3 s

output mode (L-type)

“Direct” for all aI blocks unless otherwise specified when ordered

*1

Qmax see table 1

*2

Qvmax = Qmax * 3.3 for liquids