8. GENErAL SPECIFICATIONS

8-2

IM 01R04B05-00E-E 3rd edition July 30, 2010 -00

all Rights Reserved. Copyright © 2005, Rota Yokogawa

Temperature measurement

temperature measuring range of converter :

Standard, /Lt, /Mt

: -200°C to 230°C

option /Ht

:

0°C to 350°C

accuracy:

Standard (-70°C to 150°C) : ±(0,5°C+0,005*abs(t

medium

-20°C))

option /Mt (-70°C to 230°C): ±(0,5°C+0,005*abs(t

medium

-20°C))

option /Lt (-200°C to 150°C): ±(1,0°C+0,008*abs(t

medium

-20°C))

option /Ht (0°C to 350°C) : ±(1,0°C+0,008*abs(t

medium

-20°C))

For process temperatures more than 80°C higher/lower

than ambient temperature the detector should be insulated

to maintain optimum accuracy.

Heat Tracing

Heating with heat carrier, insulation and protection housing.

the max. surface temperature at the protection housing

from inner heating is 40°C. above 150°C process tempera-

ture insulation from the manufacturer is recommended.

However

up to 230°C process temperature the customer can insulate

the detector themselves.

option /t1 : only insulation and protection

option /t2 : insulation, protection and heating line

option /t3 : like /t2 but with ventilation

Process connection for the heat carrier fluid (see table 10):

for D-type flanges : En Dn 15 Pn 40 Form B1

for a-type flanges : anSI ½ - 150 lbs.

for J-type flanges : JIS Dn15 10K

Max. pressure

: Pn 40

Protection class : IP54, install roof protected

For fluid temperatures below -70°C select option /Lt and

ask for special insulation (see also page 10).

Calibration for liquids and gases :

the RotaMaSS flowmeters are always factory calibrated

with water. Calibration Conditions:

- Water

: 22.5°C ± 12.5°C

- ambient temperature

: 22.5°C ± 12.5°C

- Process Pressure

: 1 to 2 bar abs

For gas applications please choose option /Ga.

all specifications are based on above mentioned calibration

reference conditions, a flow calibration protocol is attached

to each instrument.

Special calibrations

- Mass-/Volume flow calibration with factory certificate (option /K2):

Calibration with water at customer specified flow values

according calibration order sheet.

- Mass-/Volume flow calibration with DKD certificate

(En17025: 2005) (option /K5):

Calibration with water at customer specified flow values

according calibration order sheet.

- Density calibration with factory certificate (option /K6)

(not with /Ga):

adjustment and check with 3 different fluids, fluid temperature

influence adjustment for low ambient temperature influence

and thermal treatment for long term density measurement

stability, enhanced temperature measurement (see also

page 11).

dual Seal approval (option /dS):

- Conform with anSI/ISa-12.27.01.

- only for use with hazardous substances.

- up to anSI class 900 line pressure.

- only with FM approval option.

- For liquid application the leakage detection is realized by

software in the converter.

- For gas application options /Ga and /RD (rupture disk)

are mandatory.

- Rupture disk is only for annunciation.

NOrMAL OPErATING CONdITIONS

Ambient temperature limits
- Remote detector RCCS3 :
Standard

: -50°C to +80°C

option /Lt

: -50°C to +80°C

option /Mt

: -50°C to +80°C

option /Ht

: -50°C to +65°C

(up to 280°C medium temperature)

-50°C to +55°C

(up to 350°C medium temperature)

terminal box lower 100°C

- Remote converter RCCF311 and Integral type RCCt3:
Display operating range

: -20°C to +55°C

Electronic operating range

: -40°C to +55°C

Cold start

: above -30°C

Where meters are mounted in direct sunlight, it is recom-
mended to install a sunshade. this is particularly important in
countries with high ambient temperatures.

Ambient humidity limits : 0 to 95% R.H.

Process temperature limits
Detector :

- RCCS30 to 33

: -50°C to 150°C

- RCCS34 to 39/XR

: -70°C to 150°C

- RCCS34 to 39/XR /Mt : -70°C to 230°C

(Range 150°C – 230°C

recommended with /tx option)

- RCCS34 to 39/XR /Lt : -200°C to 150°C
- RCCS34 to 39/IR /Ht : 0°C to 350°C (only with /tx option)
For use in hazardous area -50°C is low limit.
Integral type :
- RCCt34 to 39/XR

: -50°C to 150°C

Heat carrier fluid temperature limits
(option /t2 or /t3 only for remote type RCCS30 to 39/IR)
- Standard

: 0°C to 150°C

- With option /Lt

: -200°C to 150°C

- With option /Mt

: 0°C to 230°C

- With option /Ht

: 0°C to 350°C

For fluid temperatures below -70°C select option /Lt and ask
for special insulation (see also page 10).
Process pressure limits
according to the flange ratings:
- En Pn 16

: max. 16 bar

- En Pn 40

: max. 40 bar

- En Pn 63

: max. 63 bar

- En Pn 100

: max. 100 bar

- aSME class 150

: max. 16 bar

- aSME class 300

: max. 41 bar

- aSME class 600

: max. 83 bar

- aSME class 900

: max. 124 bar

- aSME class 1500

: max. 207 bar

- JIS 10K

: max. 14 bar (1.4 MPa)

- JIS 20K

: max. 34 bar (3.4 MPa)

the RCCS30 to RCCS34 have also thread connection. For
these connections the max. allowed tube pressure is the
limitation. For all other standard process connection please
find the max. process pressure in table 9.